Product

Double Drum Autoclave

Description

Product Description

  • Economy models of our surgical double drum sterilizers are made with stainless steel sterilization chamber and painted MS from outside; while ring and stand are made of heavy MS sheet and chrome plated. The lid is also made of MS and Stainless steel sheet line is fitted inside the lid.
  • Deluxe models are made with stainless steel 304 from inner and outer and lid is also made of SS304; while ring and stand are made of chrome plated MS.
  • In order to control the pressure, our double drum surgical autoclaves are equipped with automatic pressure switch (Danfoss make). Pressure is factory set at 15 to 20psi as required.
  • A mechanical timer from 10 to 60 minute is fitted at control panel. Using this timer, operator can set time for sterilization cycle.
  • While using a high pressure sterilization machine, safety becomes important part; therefore, we equip our double drum autoclave hospital machines with spring loaded safety valve which is fitted at top of the lid. It opens if pressure rises beyond set level.
  • For heating, we use only ISI mark immersion heaters of 2 KW and 4 KW as per capacity of the autoclave. For their safety we offer low water cut-off device which is optional and fitted at extra cost.
  • We also offer some optional accessories such as digital temperature controller cum timer with END cycle buzzer, Fully automatic operation using solenoid valves, all stainless steel construction (GMP model), RS232 computer interface etc.

Product Information

Model DDA-45 DDA-75
Volume 45 Liters 75 Liters
Inner dimensions 12 x 22 Inch (Dia. x L) 16 x 24 Inch (Dia. x L)
Drum capacity 02 Nos. 11*9 inch 02 Nos. 15*12 inch
Heater ISI mark Immersion heater 2 KW ISI mark Immersion heater 4 KW
MOC Exterior Painted MS
Stand and ring Chrome plated MS
Sterilization chamber SS 304
Lid Painted MS w/ SS inner line
Locking system Radial locking mechanism w/ pedal lifting device
Pressure 15 to 20 psi
Temperature 121°C to 125°C
Pressure release Manual release value
Pressure display Dial pressure gauge
Pressure control Automatic pressure control switch
Gasket High temperature joint less silicone rubber gasket
Timer Mechanical timer adjustable up to 60 minutes
Safety Spring loaded safety valve on lid
Power supply 220 Volts 50Hz
Test confirmation Hydraulic tested at 2.5 times of working pressure

Reviews

There are no reviews yet.

Be the first to review “Double Drum Autoclave”

Your email address will not be published. Required fields are marked *